TRANCE(TRiggeredAngiographyNonContrastEnhanced)非造影剂增强门控血管造影
;
TRANCE(TRiggeredAngiographyNonContrastEnhanced)isatechniqueforimagingarteriesandveinswithoutcontrastagents. ThisExamCardisintendedforthelowerlegs.SENSE2isusedforspeedup.
TRANCEprotocolsare3DTSEsequenceswithcardiactriggering.Visualizingarteriesonlyrequirestwoseparatescanswithtriggeringindifferentcardiacphases.Insystolearterialbloodisflowingfast.Thiscausesdephasingofsignalandleadstoflowvoids,soarterieswillbeblackwithsystolictriggering.Indiastolebloodflowinarteriesisslow.Thesignalwillnotdephase,soarterieswillbebrightondiastolicscans.Subtractingthesescansresultsina3Ddatasetwitharteriesonly.
Toobtainveinsonly,a3DTSEprotocolwithSTIRisscannedwithsystolictriggering.STIRgivesextrabackgroundsuppressionasfatandboneswillalsobesuppressed.Withsystolictriggeringthearteriesareblack.Resultisadatasetwithveinsonly,nosubtractionneeded.
TRANCE(TRiggeredAngiographyNonContrastEnhanced) 非造影剂增强门控血管造影技术是一种无需注射造影剂的高对比MR血管成像技术。它利用心电触发的3DTSE序列,在同一心动周期的不同时相进行两次扫描以获得动脉影像。在收缩期的时候,动脉内的血液由于快速流动产生失相位而导致信号丢失,从而使得在收缩期触发采集的动脉呈低信号。在舒张期时,动脉内的血流因为流速相对缓慢而不会产生失相位,动脉呈高信号。将舒张期和收缩期采集的影像进行减影后,即得到动脉的3D影像。
如果单独在舒张期进行门控触发的3DTSE序列采集,即可获得静脉影像,同时可结合STIR序列以抑制脂肪和骨信号,从而获得较好的背景抑制效果。此时动脉呈低信号。
为了确保在舒张期和收缩期进行门控触发,必须在有关采集前先进行QFlow的扫描,通过血流动力学的分析以确定正确的触发延迟时间。
章伟敏(来源于飞利浦公司)